BRONKHORST HIGH-TECH B.V.
2.13.5 Solenoid Operated shut off valves NC and NO
- Removal
Ensurethatthesystemisdepressurisedbeforetakingshutoff
valveapart.
Thepowermustbeoffwhenremovingtheelectricalconnectors.
Removethecoilbyreleasingtheclipontopofthecoilforthe
normallyclosed,andthenutforthenormallyopened.
Loosenthe2boltswithhexagonheadcounterclockwiseto
releasethevalve.
- Replacement
ReplacethegasketsAS‐007andAS‐017,Viton,EPDM
orKalrez.
Allwettedpartshavetobeabsolutelyclean,don’tuseoilor
greaseandavoiddust!
Replacetheshutoffvalveandtightentheboltwithhexagon
headbyturningclockwise,tightencrosswisewithM30,9Nm.
Replacethecoilandreplacetheclipontopofthecoilnormally
closed,andthenutforthenormallyopened.
Replacetheelectricalconnector.
Nowtheshutoffvalvehastobetestedforleakageandfunction.
Solenoid
valve coil
VDR
LED
Connector
1
2
Availableelectricshutoffvalves
Modelnumber Description
EV‐02‐NC‐V(iton) Kv=0,065 MaxdeltaP3bar
EV‐02‐NC‐E(PDM) Kv=0,065 MaxdeltaP3bar
EV‐02‐NC‐K(alrez) Kv=0,065 MaxdeltaP3bar
EV‐02‐NO‐V(iton) Kv=0,065 MaxdeltaP3bar
Hook‐upsolenoidoperatedvalve.
Supplyvoltage+24Vdc
Powerconsumption3Watt.
2.13.6 Toggle valves
- Removal
Ensurethatthesystemispressurelessbeforetaking
thetogglevalveapart.
Loosenthe4boltswithhexagonheadcounterclockwise
toreleasethetogglevalve.
- Replacement
ReplacethegasketsAS‐007andAS‐017,Viton,
EPDMorKalrez.
Allwettedpartshavetobeabsolutelyclean,don’tuse
oilorgreaseandavoiddust!
Replacethetogglevalvecentrebeforetightenthebolt
withhexagonheadbyturningclockwise,tightencrosswise
withM30,9Nm.
Nowthetogglevalvehastobetestedforleakageandfunction.
AvailableTogglevalves
Modelnumber Description
TV‐01‐V Kv=0,09 ToggleValvewithVitonseals
TV‐01‐E Kv=0,09 ToggleValvewithEPDMseals
TV‐01‐K Kv=0,09 ToggleValvewithKalrezseals
9.17.057 page 13