BRONKHORST HIGH-TECH B.V.
ThecontrolvalveusedintheFLOW‐SMSseriesisastandard,direct
operatedsolenoidvalve(normallyopenedornormallyclosed),
operatedthroughthePID‐controlfunctiononthepc‐boardof the
floworpressuremeter.Thediameteroftheorificeunderthe
plungerisoptimisedforthecustomer’sspecification.
9.17.057 page 12
InstructionsforreplacingaPressureMeter(PM)or
PressureController(PC)module:
- Removal
Ensurethatthesystemisdepressurisedbeforetaking
Pressuremoduleapart.
Thepowermustbeoffwhenremovingtheelectricalconnectors.
Loosenthe4boltswithhexagonheadcounterclockwise to
releasethePressuremodule.
- Replacement
Replacethegaskets2xAS‐108,Viton,EPDM
orKalrez.
Allwettedpartshavetobeabsolutelyclean,don’tuseoilor
greaseandavoiddust!
ReplacethePressuremoduleandtightentheM4boltswithhexagon
headbyturningclockwise,tightencrosswisewith2,0Nm.
Replacetheelectricalconnector.
NowthePressuremodulehastobetestedforleakageand
function.
AvailablePressureController/PressureMetermodules
PressureModule Description
Modelnumber Function Smallestrange Highestrange
PS‐502C Meter 2..100mbar 0,2…10bar
PS‐602CV‐NC Controllerforward(NC) 5..100mbar 0,5…10bar
PS‐702CV‐NC Controllerback(NC) 20…100mbar 2…10bar
Pressurerange(FS) Indentificationinmodelkey
(bar) Absolute Relative
0,1‐0,35
350A 350R
0,35‐1,1
1k1A 1k1R
1,1‐6,0
6k0A 6k1R
6,0–21
21kA 21kR
PleasekeepinmindthatthepressureratingforFLOW‐SMSis10baronly.
Inpressurecontrolsystemsthesystemwidelydeterminestheresponsebehaviourofthecontrolloop.
Duringtestingthecustomersystemissimulatedascloselyaspossible.Insomecaseshowever
readjustmentisneededforoptimumperformanceunderactualconditions.
Comentários a estes Manuais