Page 42 Operational instructions for digital multibus instruments 9.17.023
11.2 LED’S INDICATIONS
11.2.1 LED indications mode (no switch used)
Power-off or program not running
Normal running/operation mode
Initialization mode (Init reset = 73)
For MBC3 type: no bus communication, safe state active.
Special function mode
Instrument is busy performing any special function. E.g. auto-zero or self-test
0.1 sec off
Special mode, see specific field bus for more details
flash
2.0 sec off
Special mode, see specific field bus for more details
Node occupied: Re-install instrument
No data-exchange between master and slave Automatic recovery
Data is received or transmitted
Minor communication error
Instrument is not in OP mode (see EtherCAT manual for details)
flash
0.2 sec off
An error occurred of minor importance.
It would be wise to investigate the cause of this.
You are still able to work with your instrument.
See specific field bus for more details
Waiting for communication
0.1 sec off
See specific field bus for more details
A requested parameter is not available.
See troubleshoot in Profibus manual.
For special service purpose only
Serious communication error; manual intervention needed
Error detected in EtherCAT configuration (see EtherCAT manual
for details)
Critical error message. A serious error occurred in the instrument.
Instrument needs service before further using.
Wink Mode Green Red Green Red turn by turn
wink
0.2 sec off
By a command send via FLOW-BUS the instrument can “wink” with Led’s to indicate its
position in a (large) system
Alarm indication: minimum alarm, limit/maximum alarm; power-up alarm or limit exceeded
or batch reached.
Switch-released, selected action started