Details and meaning can be found in document 9.17.023 Operation instructions digital instruments.
This document can be found at:
http://www.bronkhorst.com/en/downloads/instruction_manuals/
• Access indicates whether parameter can be Read and/or Written.
• When a byte parameter is read, the upper 8-bits of the Modbus register will be 0. When a byte
parameter is written, the upper 8-bits must be set to 0.
• Long integer parameters have a length of 4 bytes and are mapped on two consecutive Modbus
registers. The first register contains bit 32-16, the second register contains bit 15-0.
• Floating point parameters have a length of 4 bytes and are mapped on two consecutive Modbus
registers. Floats are in single precision IEEE format (1 sign bit, 8 bits exponent and 23 bits fraction).
The first register contains bit 32-16, the second register contains bit 15-0.
• String parameters can have a length of maximal 16 bytes and can take up to 8 Modbus registers
where each register contains two characters (bytes). The upper byte of the first register contains
the first character of the string. When writing strings, the write action should always start from the
first register as a complete block (it is not possible to write a part of a string). If the string is shorter
than the specified maximum length the string should be terminated with an 0.
• Parameters Temperature, Counter value and Counter limit can be found in the parameter table as
an unsigned integer variant and as a floating point variant. Only the floating point variant supports
the full parameter range and resolution.
PARAMETER NAME
PARAMETER
TYPE
ACCESS
PDU ADDRESS REGISTER NUMBER
REMARK
Hex Dec Hex Dec
64841..64842
Page 20 Modbus interface 9.17.035