2.10 System purging
Ifexplosive gases are tobeused,purgethe process with inertdrygaslikeNitrogen, Argon etc. forat
least 30 minutes. Complete purging is also required to remove such gases from the system before
exposingthesystemtoair.
2.11 Seals
Thesealsfittedarecompatiblewiththepublishedlistofgases,atthestatedconditions.
2.12 Equipment storage
Theequipmentshouldbestoredinitsoriginalpackinginacupboard,warehouseorsimilar.Careshould
betakentonotsubjecttheequipmenttoexcessivetemperaturesorhumidity.
2.13 Electromagnetic compatibility
TheMASSVIEW®seriescarriestheCEmark.
ThereforeithastocomplywiththeEMCrequirementsasarevalidfortheseinstruments.
HowevercompliancewiththeEMCrequirementsisnotpossiblewithouttheuseofpropercablesand
connectorassemblies.Donotuseunshieldedwireterminals.